DQF-COSY with WATERGATE

Description Gradient-enhanced phase-sensitive DQF COSY with WATERGATE solvent suppression incorporated into a refocusing delay. Employs magic angle field gradients.
References L. A. Trimble and M. A. Bernstein, J. Magn. Reson. B, 105, 67-72 (1994).

P. C. M. van Zijl et al, J. Magn. Reson. A, 113, 265-270 (1995).

Pulse Sequence cosygpw.bf
Parameter Set cosygpw.bf
Parameters to be Adjusted p1, d1, ns, sp1, phcor11

Instructions

  1. Acquire a 1D data set to determine SW. Calibrate p1 (1H 90° pulse at pl1=0, 9 to 13 µs).

  2. Create a 2D experiment and read in parameter set.

  3. Set p1. Adjust d1, 1 sw = 2 sw and ns as required for your sample.

  4. In gs mode adjust sp1 and phcor11 iteratively until optimal solvent suppression is attained.